MSA Ultima X Replacement Sensor Guide: A-ULTX-SENS Gas Codes and PCB Options

How to Decode Replacement Sensors, Match Monitor Construction, and Identify PCB Configurations

Upgrading or replacing your MSA Ultima X gas detector sensor requires matching the correct A-ULTX-SENS gas code with your transmitter’s PCB architecture. This guide decodes sensor part numbers, pin configurations, and board options to ensure seamless, reliable safety compliance.Most people landing on this page are in one of three situations. You may have an Ultima X displaying SENSOR WARNING or CHANGE SENSOR. You may be trying to decode a damaged sensor label. Or you may be repairing a transmitter that is reporting a main-board fault. This guide covers all three, with the A-ULTX-SENS ordering tables first and the PCB and fault-message references behind them.

MSA has discontinued the Ultima X Series and identifies the Ultima X5000 as the current alternative. That does not automatically make every working Ultima X point a migration project. If the transmitter is healthy and the exact approved sensor is available, a sensor replacement can still be the cleanest way to support a standardized installed base..

The practical rule: Use the gas code to narrow the search, but use the complete A-ULTX-SENS or A-ULTX-PCB label to place the order. Gas, range, construction, approval, wiring, and configured options all matter.

1. How Ultima X Replacement Part Numbers Work

The official product family is MSA Ultima X. You may see A-ULTIMAX in complete-monitor catalog numbers, while the replacement sensor and circuit-board families use the shorter A-ULTX prefix. The three numbers below solve different maintenance problems:

Part-number family What it identifies What must match
A-ULTX-SENS Configured smart sensor module Gas, full-scale range, construction / approval, and final configured suffix
A-ULTX-PCB Configured main printed circuit board Monitor family, output / wiring, HART or Modbus, LEDs / relays, language, software, and custom options
A-ULTIMAX Complete configured monitor assembly The entire transmitter model and every selected option

A body, guard, calibration cap, or flow cap is a separate physical accessory. This article focuses on the two most common sourcing questions: the replacement sensor module and the configured main PCB.

2. How to Read an A-ULTX-SENS Part Number

The supplied ordering-sheet material uses the format A-ULTX-SENS-[gas code]-[enclosure code]-[custom code]. Some catalog records compress the final characters, so copy the number exactly as printed.

Position Example Meaning in A-ULTX-SENS-16-0-0
Prefix A-ULTX-SENS Ultima X configured replacement sensor
Gas / range code 16 Hydrogen sulfide, 0-50 ppm
Construction code 0 General-purpose plastic construction
Custom / configured code 0 Standard option shown in this example

A-ULTX-SENS-16-0-0 and A-ULTX-SENS-16-1-0 share the same gas and range. They are still different parts because the construction code changes from general-purpose plastic to explosion-proof stainless.

3. Complete Gas and Range Code Tables

The three tables below preserve the gas codes and ranges from the MSA A-ULTX-SENS ATO ordering sheet.

Toxic gas and oxygen sensor codes

Code Gas Range ATO ordering note
11 Carbon monoxide 0-100 ppm Standard
12 Carbon monoxide 0-500 ppm Standard
57 Carbon monoxide 0-1000 ppm Standard
13 Oxygen 0-10% Standard
14 Oxygen 0-25% Standard
55 Oxygen, solvent-tolerant 0-25% XP stainless only
62 Oxygen, solvent- and CO2-tolerant 0-25% XP stainless only
63 Oxygen, low range Per specification XP stainless only
64 Oxygen, low range and solvent-tolerant Per specification XP stainless only
15 Hydrogen sulfide 0-10 ppm Standard
16 Hydrogen sulfide 0-50 ppm Standard
17 Hydrogen sulfide 0-100 ppm Standard
25 Hydrogen sulfide 0-500 ppm Standard in the supplied ATO transcription
18 Chlorine 0-5 ppm GP stainless, plastic, ATEX, or UL Division 2
28 Chlorine 0-10 ppm GP stainless, plastic, ATEX, or UL Division 2
61 Chlorine 0-20 ppm GP stainless, plastic, ATEX, or UL Division 2
20 Nitric oxide 0-100 ppm Standard
22 Hydrogen cyanide 0-50 ppm Standard
23 Hydrogen chloride 0-50 ppm GP stainless, plastic, ATEX, or UL Division 2
24 Chlorine dioxide 0-3 ppm GP stainless, plastic, ATEX, or UL Division 2
26 Hydrogen fluoride 0-10 ppm GP stainless, plastic, ATEX, or UL Division 2
70 Sulfur dioxide 0-25 ppm IS barrier with ATEX, or UL Division 2
71 Sulfur dioxide 0-100 ppm IS barrier with ATEX, or UL Division 2
72 Nitrogen dioxide 0-10 ppm IS barrier with ATEX, or UL Division 2

Catalytic combustible and infrared sensor codes

Code Gas / technology Range or calibration basis ATO ordering note
31 Combustible, catalytic 0-100% LEL, natural gas and hydrogen Standard
32 Combustible, catalytic 0-100% LEL, petroleum vapors Standard
33 Combustible, catalytic 0-100% LEL, solvents Standard
51 Combustible, catalytic, ATEX 0-100% LEL, natural gas and hydrogen ATEX calibration basis
52 Combustible, catalytic, ATEX 0-100% LEL, petroleum vapors ATEX calibration basis
53 Combustible, catalytic, ATEX 0-100% LEL, solvents ATEX calibration basis
38 Combustible, infrared 0-100% LEL, methane Custom order in the supplied ATO transcription
39 Combustible, infrared 0-100% LEL, propane / non-methane Custom order in the supplied ATO transcription
58 Combustible, infrared, ATEX 0-100% LEL, methane Standard in the supplied ATO transcription
59 Combustible, infrared, ATEX 0-100% LEL, non-methane Standard in the supplied ATO transcription
34 Acetylene, infrared 0-2.5% by volume Custom order
35 Carbon dioxide, infrared 0-5% by volume Standard in the supplied ATO transcription
36 Carbon dioxide, infrared 0-2% by volume Standard in the supplied ATO transcription
37 Carbon dioxide, infrared 0-5% by volume A second 0-5% selection; the supplied material does not explain the distinction

Specialty toxic gas sensor codes

Code Gas Range ATO ordering note
41 Phosphine 0-2 ppm Standard
42 Arsine 0-2 ppm Standard
43 Silane 0-25 ppm Standard
45 Diborane 0-50 ppm Standard
46 Bromine 0-5 ppm GP stainless, plastic, ATEX, or UL Division 2
47 Fluorine 0-10 ppm in the supplied ATO transcription Manual cross-check differs; see revision note below
48 Ammonia 0-100 ppm GP stainless, plastic, ATEX, or UL Division 2
54 Ammonia 0-1000 ppm GP stainless, plastic, ATEX, or UL Division 2
49 Hydrogen 0-1000 ppm Standard
50 Ethylene oxide 0-10 ppm GP stainless, plastic, ATEX, or UL Division 2

What the instruction manual adds or changes

The ATO sheet is the better catalog decoder; the instruction manual is the better maintenance and troubleshooting source.

Code / group Manual entry Why it matters when ordering
44 Germane, 0-3 ppm Appears in the instruction-manual replacement table but was omitted from the supplied ATO transcription.
47 Fluorine The supplied ATO transcription shows 0-10 ppm; the instruction-manual replacement table shows 0-5 ppm. Match the printed sensor label and document revision.
63 Low-range oxygen The supplied ATO transcription says “per specification”; the instruction-manual replacement table shows 0-25% for the listed Model E construction.
64 Low-range, solvent-tolerant oxygen The supplied ATO transcription says “per specification”; the instruction-manual replacement table shows 0-25% for the listed Model E construction.
ATO-only selections Codes 25, 26, 28, 34-37, 51-53, and 58-59 These appear in the supplied ATO transcription but are not all reproduced in the manual’s standard replacement table. That does not make them interchangeable or invalid; it makes the full label and approval essential.

Revision rule: For a discontinued platform with multiple approval variants, the full installed label outranks a generic two-digit code table. When a range or suffix differs between documents, send the label to TCI Supply.

4. Match the Construction and Enclosure Code

Code Construction Practical interpretation
0 General-purpose plastic Commonly associated with Model A / Ultima XA configurations
1 316 stainless explosion-proof Commonly associated with Model E / Ultima XE; the manual also lists selected XIR sensors with this suffix
6 General-purpose stainless Model G construction in the instruction-manual replacement table
2 through 12 IR, ATEX, and regional stainless constructions The supplied ATO transcription groups these as special construction values. Copy the exact installed number; do not infer the suffix from the gas alone.

The manual replacement table makes the common pattern easy to see: a gas code may be offered with -0 for general-purpose plastic, -1 for the Model E explosion-proof construction, or -6 for general-purpose stainless. Not every gas is available in every construction. Also check the approval label. A sensor that appears to thread into the housing is not automatically the correct replacement.

5. Common A-ULTX-SENS Examples

Example part number Gas and range Construction
A-ULTX-SENS-11-0-0 Carbon monoxide, 0-100 ppm General-purpose plastic
A-ULTX-SENS-14-0-0 Oxygen, 0-25% General-purpose plastic
A-ULTX-SENS-16-0-0 Hydrogen sulfide, 0-50 ppm General-purpose plastic
A-ULTX-SENS-16-1-0 Hydrogen sulfide, 0-50 ppm Explosion-proof stainless
A-ULTX-SENS-31-0-0 Catalytic combustible, natural gas / hydrogen basis General-purpose plastic
A-ULTX-SENS-38-1-0 Infrared combustible, methane basis Listed Model E / XIR construction
A-ULTX-SENS-57-6-0 Carbon monoxide, 0-1000 ppm General-purpose stainless

6. How to Identify an A-ULTX-PCB Replacement Board

A sensor is selected primarily by gas, range, and construction. A PCB is selected by the monitor family and its electrical and option configuration. The supplied catalog defines six positions after A-ULTX-PCB:

Position Options in the supplied ATO decoder
Model type A = general-purpose plastic (XA); E = explosion-proof 316 stainless (XE); T = plastic, no display (XT); L = stainless, no display (XL)
Sensor output 1 or B = 4-20 mA two-wire toxic / O2 with no LEDs or relays; 2 = two-wire HART; 3 or E = three-wire with LEDs or relays; 4 or H = three-wire HART; M = Modbus
LEDs and relays 0 = none; 1 = LEDs only; 2 = relays only; 3 = LEDs and relays
Language 0 = English; S = Spanish
Software 0 = North American; 1 = European; 3 = custom horn
Custom features 0 = none; 1 = internal 12 VDC supply; 2 = internal 24 VDC supply; C = custom

The instruction manual provides a useful cross-check for the core A and E numeric board prefixes:

Core board prefix Monitor construction Output configuration
A-ULTX-PCB-A-1 XA / general-purpose plastic 2-wire, 4-20 mA
A-ULTX-PCB-A-2 XA / general-purpose plastic 2-wire, 4-20 mA with HART
A-ULTX-PCB-A-3 XA / general-purpose plastic 3-wire, 4-20 mA
A-ULTX-PCB-A-4 XA / general-purpose plastic 3-wire, 4-20 mA with HART
A-ULTX-PCB-E-1 XE / XIR explosion-proof 2-wire, 4-20 mA
A-ULTX-PCB-E-2 XE / XIR explosion-proof 2-wire, 4-20 mA with HART
A-ULTX-PCB-E-3 XE / XIR explosion-proof 3-wire, 4-20 mA
A-ULTX-PCB-E-4 XE / XIR explosion-proof 3-wire, 4-20 mA with HART

7. Sensor or PCB? Use the Ultima X Fault Message

The ATO sheet identifies parts; it is not a troubleshooting guide. For diagnosis, the MSA instruction manual gives a better starting point because its displayed messages distinguish sensor-module faults, main-PCBA faults, and calibration failures.

Display message What it indicates Practical next step
SENSOR WARNING The sensor is approaching end of life Plan the correct A-ULTX-SENS replacement. The warning itself does not change the analog output.
CHANGE SENSOR The sensor is at end of life Replace the sensor module and perform the required calibration before returning the point to service.
SENSOR MISSING The monitor has lost communication with the sensor Check the sensor connection; connect or replace the correct sensor module.
SNSR FLASH FAULT Sensor-module program memory is invalid Replace the sensor module.
SNSR RAM FAULT Sensor-module memory fault Replace the sensor module.
SNSR DATA FAULT The sensor data sheet is invalid Reset through the controller; replace the sensor if the fault remains.
INVALID SENSOR The attached sensor is not compatible with the instrument Install the correct gas, range, construction, and configured sensor type.
MN FLASH / MN RAM / MN EEPROM FAULT Main-PCBA memory fault Replace the main PC board.
MN SUPPLY FAULT Main-PCBA supply is outside its permitted range Check the sensor wiring; replace the main PC board if the fault persists.
ZERO CAL FAULT The monitor could not complete zero calibration Check zero gas / clean air and the gas path, then repeat calibration before condemning hardware.
SPAN CAL FAULT The monitor could not complete span calibration Check the calibration gas, regulator, tubing, flow, and blockage before replacing a part.
LOW SIGNAL Low infrared signal Follow the XIR optics-cleaning procedure or replace the sensor as directed by the manual.
CHECK CAL Calibration should be verified Perform the site-required bump test or calibration.

This is why “replace the sensor first whenever you are unsure” is not a sound universal rule. CHANGE SENSOR points directly to the sensor. MN EEPROM FAULT points to the main board. SPAN CAL FAULT can be caused by calibration gas, regulator, tubing, flow, blockage, or the sensor, so the gas path should be checked.

8. Sensor Replacement and Calibration Notes

The sensor is the routine maintenance item and has a finite life, but the source ordering sheet does not support one universal replacement interval for every gas and technology. A warranty period is also not the same thing as a guaranteed service life. Use the calibration program, the sensor’s ability to complete calibration, and SENSOR WARNING or CHANGE SENSOR to plan replacement.

  • Identify the required sensor from the complete A-ULTX-SENS code on the interior label.
  • Replacing a sensor with the same gas type preserves the existing alarm setpoints and relay functions.
  • Replacing it with a different gas type loads the new sensor’s default alarm setpoints and upscale / downscale relay behavior. Review the PLC or DCS scaling, alarm logic, labels, and procedures before placing a different gas in service.
  • Perform the calibration with the new sensor before returning the point to service.
  • The manual describes a 60-second Sensor Swap Delay, but the permitted replacement procedure depends on the exact model, approval, hazardous-area classification, and site procedure. Do not treat SafeSwap as a blanket hot-swap instruction.
  • Use genuine MSA replacement parts.

Life-safety equipment note: This article is a part-identification and sourcing guide. It does not replace the MSA instruction manual, approval drawings, or your site gas-detection, lockout, hazardous-area, calibration, and commissioning procedures.

9. Keep the Ultima X Running or Move to the X5000?

MSA identifies the Ultima X5000 as the current alternative. The practical decision depends on whether the problem is a normal sensor replacement or a broader transmitter-lifecycle issue.

Keep supporting the Ultima X Evaluate an Ultima X5000 retrofit
The transmitter is healthy and only its sensor is due for replacement. Main-board faults are recurring or the exact configured board is difficult to source.
The site has a standardized fleet and already stocks matching Ultima X spares. The project is already opening the loop, changing the enclosure, or modernizing the gas-detection system.
The complete A-ULTX-SENS number is known and the correct approved module is available. You want a current MSA platform and can revalidate power, wire gauge, approvals, alarm logic, and commissioning.
Keeping the existing point avoids an unnecessary change to a proven installation. The identical mounting holes and conduit-entry geometry can reduce mechanical retrofit work, subject to the current MSA installation requirements.

MSA designed the X5000 for retrofit applications. Its mounting holes and conduit-entry locations match the Ultima X, and existing wiring can be reused when the wire gauge and available power meet the current X5000 requirements.

The X5000 uses the A-5K-SENS replacement-sensor family. Legacy A-ULTX-SENS spares do not transfer. A sensible installed-base plan can support healthy Ultima X points with exact replacement sensors while moving selected failed or strategic points to the X5000 during planned work.

10. What to Send TCI Supply for an Accurate Quote

The fastest quote is built from labels, not a typed description such as “MSA H2S sensor.” Send the following:

  • A clear photo of the full transmitter model and approval label.
  • A clear photo of the interior sensor label showing the complete A-ULTX-SENS code.
  • The gas and printed full-scale range.
  • Plastic or stainless construction, and whether the point is general-purpose, explosion-proof, ATEX, or another approved configuration.
  • For a PCB: the full A-ULTX-PCB label, supply voltage, 2- or 3-wire output, HART or Modbus requirement, local display, and relay use.
  • Quantity, ship-to country, required date, and any required approval or certificate.

If the label is damaged, send several close-up photos plus one wider photo showing the monitor body, sensor assembly, conduit entries, and installation. That usually provides more useful identification detail than a partial typed number.

11. Source MSA Ultima X Parts From TCI Supply

TCI Supply sources discontinued and hard-to-find industrial control and safety products, including configured Ultima X sensors and boards. Browse the MSA Safety catalog, review the verified A-ULTX-SENS-16-0-0 H2S sensor, or contact TCI Supply with label photos. You can also call 1-800-755-2883.

Reference documents: A-ULTX-SENS ATO ordering sheet and Ultima X Series instruction manual.

Frequently Asked Questions

What does A-ULTX-SENS mean?
A-ULTX-SENS is the configured replacement-sensor family used by the MSA Ultima X platform. The complete number includes the gas and range code, the construction code, and a final configured option. The full interior label is the orderable reference.
Can I use the same gas code with a different construction code?
No. The gas code is only one position. A code 16 H2S sensor for a general-purpose plastic monitor is not the same part as code 16 for an explosion-proof stainless monitor. Match the complete A-ULTX-SENS number and approval label.
Can I change the gas type by installing a different smart sensor?
MSA documents that a same-gas replacement preserves existing alarm and relay settings, while a different gas type loads the new sensor’s defaults. A gas change also changes what the PLC, DCS, relays, labels, and procedures must represent.
How do I tell whether I need a sensor or a PCB?
Start with the displayed fault message. SENSOR WARNING, CHANGE SENSOR, SENSOR MISSING, and SNSR memory faults point toward the sensor module. MN FLASH, MN RAM, and MN EEPROM faults point toward the main PC board. Calibration faults require gas and flow checks before hardware is condemned.
How long does an Ultima X sensor last?
There is no single service-life number that applies to every gas, sensor technology, exposure history, and environment. Use your calibration schedule, calibration performance, and the monitor’s SENSOR WARNING or CHANGE SENSOR messages to plan replacement.
Are Ultima X replacement parts still available?
Availability varies by gas, range, construction, approval, and configured suffix. Common modules and selected boards are in stock, while uncommon configurations may require a sourcing search.
Is the Ultima X5000 a drop-in replacement for the Ultima X?
It is retrofit-friendly, not a sensor-level drop-in replacement. MSA documents identical mounting holes and conduit-entry geometry and allows reuse of existing wiring when wire gauge and power requirements are met. The X5000 uses the A-5K-SENS family, so A-ULTX-SENS spares do not transfer.